Performance Characte ristics | |
Output signal | 700 ± 250 nA / ppm |
Typical Baseline Range (pure air) | ±2 ppm CO equivalent |
T90 Response Time | < 30 seconds |
Measurement Range | 0 - 100 ppm |
Maximum Overload | 500 ppm |
Linearity | Linear |
Repeatability | < ±2% H2S equivalent |
Recommended Load Resistor | 10 ohms |
Resolution (Electronics dependent) | < 0.1 ppm typical |
Environmental | |
Temperature Range Continuous | -30oC to +50oC |
Pressure Range | 800 to 1200 mbar |
Operating Humidity Range | 15% to 90% RH |
Lifetime | |
Long Term Output Drift | < 5% per annum |
Recommended Storage Temp | 0oC to 20oC |
Expected Operating Life | > 24 months in air |
Standard Warranty | 24 months from date of dispatch |
Intrinsic Safety Da ta | |
Maximum at 2000 ppm | 0.3 mA |
Maximum o/c Voltage | 1.3 V |
Maximum s/c Current | <1.0 A |
Cross -Sensitivity Data
Cross -Sensitivity Dat
GAS | CONC. | GS+4H2S |
Carbon Monoxide | 100 ppm | <2 ppm |
Sulphur dioxide | 20 ppm | 0 ppm |
Nitrogen Dioxide | 5 ppm | <0.5 ppm |
Nitric Oxide | 50 ppm | <0.5 ppm |
Ammonia | 50 ppm | 0 ppm |
Chlorine | 15 ppm | 0 ppm |
Ethylene | 100 ppm | 0 ppm |
Carbon Dioxide | 5000 ppm | 0 ppm |
Important Note: The values above are typical values and should not be used as a basis for cross calibration. Cross sensitivities may not be linear and should not be scaled either. Above data based on gassing for 5 minutes using CCR Scientific test equipment. Should be noted some cross interference break through will occur if gas is applied for a longer period of time.
Poisoning:
CCR Scientific sensors are designed to operate in a wide range of harsh environments and conditions. However, it is important that exposure to high concentrations of solvent vapors is avoided, both during storage, fitting into instrument and operation.